Application Note: Moisture Determination of Cement

Precisa moisture content analysis XM60/EM120 Halogen

Moisture Content of Cement

This information is intended as a guideline for development of a moisture analysis method and parameter setting. Materials vary throughout the world and we will be more than happy to test customers’ individual samples.

Description:Cement, fine powder
Utensils:Aluminium pan
Preparation:Spread thinly on the pan
Sample weight (+/- 10%):2g
Temperature program:Standard
Temperature:105 oC
Standby Temperature:100 oC
Stop-Criteria:Autostop 2 D/ 60 s
Moisture content: (average of 4 measurements)1.41%
Standard Deviation (+/-):0.02%
Average analysis time:3 min

Download our Moisture Content Application Note

You can download the complete application note as a PDF: MA0805 – Cement Data Sheet pdf

Precisa Moisture Analyser XM60 and Precisa Moisture Analyser EM120

Precisa Moisture Analyser XM60/EM120

If you would like to find out more about our Series 330 XM or our Series 365EM for moisture analysis, then please don’t hesitate to contact our sales and marketing team. They’ll be pleased to discuss your requirements.