Setting USB serial port

A PC can be connected directly to the USB DEVICE connector on the back side of the balance.
For data transmission, the interface parameters of the balance must match with those of the PC.

Data reception

A list of possible remote control commands is provided at the end of this help text.

Data sending

Mode Protocol

In mode Protocol, the protocol defined for the application is output automatically or with the [PRINT]-key, depending on the setting.

Mode Measurement transfer

In mode Measurement transfer by pressing the soft button Transfer or the [PRINT]-key respectively (depending on the trigger setting), the current weight value is transferred according the set format of the Measurement transfer. If trigger is set on transfer button, the soft button [Transfer] must be activated in the Action bar.

Note, to ensure to activate on the Action bar the soft button [Transfer] in the mode Measurement transfer.

 

Remote control commands

Each remote command has to be closed with CR LF which corresponds to pushing the [Enter] key. Commands can be acknowledged using ACK1.

Command specification

ACK0       : Acknowledge sending mode Off

ACK1       : Acknowledge sending mode On

BINFO      : Print out Balance information

CAL        : Start adjustment

D......    : Write to weight display (right-aligned)

DN         : Reset weight display

@......    : Write to secondary display (right-aligned)

@N         : Reset secondary display

FWHISPRINT : Print out update history

PDT        : Print out date and time

PRT        : Start printing (corresponds to pressing the [PRINT] key)

P0         : Individually print each value (unstable or stable)

P1         : Individually print value (stable)

P2 lll     : Print after change of load (lll = min. load change necessary

             in g or in percent)

             (Examples: P2 12.3 or P2 50%)

P3         : Print after each display refresh (refresh rate)

P4 ttt.t   : Print on time basis ttt.t in s, from 0.1s to 600.0s 

             (Example: P4 2.0 individual print out every 2s without waiting

             until stable)

SDTddmmyyhhmmss: Set date/time if Date format = ddmmyy

SDTmmddyyhhmmss: Set date/time if Date format = mmddyy

SOUND      : Beep tone

SOUNDI     : Info tone

SOUNDW     : Warn tone

T (ttt)    : Tare or set tare to a specific value

Z0         : Auto-Zero mode Off

Z1         : Auto-Zero mode On

ZERO       : Zero balance (provided weight is within the zero-position

             range)

 

Application Counting:       

Rnnn       : Set current weight = nnn pieces

             (corresponds to soft button "Set Pcs")

REFrrr     : Set reference weight = rrr for 1 piece

             (corresponds to soft button "Set piece weight")

RC         : Clear pieces unit (corresponds to soft button "Clr Pcs")

 

 

SICS - Command list

Level 0

@          : Reset

I0         : Send list of available SICS commands

I2         : Send device data

I3         : Send firmware version

I4         : Send serial number

S          : Send stable weight value

SI         : Send weight value immediately (only for non-approved balances)

SIR        : Send weight value immediately and repeat (approved balances: only stable weights)

Z          : Zero

ZI         : Zero immediately (only for non-approved balances)

 

Level 1

D          : Write text to display

DW         : Clear text

T          : Tare

TA         : Tare weight value

TAC        : Clear tare weight value

TI         : Tare immediately (only for non-approved balances)

 

Level 2

C2         : Start adjustment with external weight

             (only for non-approved balances or approved balances of class I)

C3         : Start adjustment with internal weight

SNR        : Send stable weight value and repeat on stable weight change